Skip to content

MYAH PRO

Topcon · myopia management

Topcon MYAH PRO : MYAH PRO

From the first pair of glasses to the first implant.

Topcon MYAH PRO – myopia management

Highlights.

Myopia

Axial length, on growth curves.

Childhood myopia is rising in Africa. Axial length, measured without contact or topical anaesthesia, is plotted visit after visit on age-based reference curves. When it moves up a curve, you see it, and parents understand it.

14–40 mm
axial length, to 0.01 mm
4.7%
childhood myopia in Africa (Kobia-Acquah 2022)
Age →Axial length →Visit 1Visit 2Visit 3It moves up a curve.
Schematic. The software's reference curves come from European populations; they help follow-up, they do not make a diagnosis.

One patient, one device

Six exams. One seat.

The patient stays at one device and nothing is re-typed. Your technician takes the measurements; you see the patient once, with the full work-up.

6 + 1
core exams, optional module
DICOM
worklist, XML, Harmony
  1. Objective refraction
  2. Topography
  3. Aberrometry
  4. Pupillometry
  5. Axial length
  6. Ocular surface
  7. Biometry and IOL*Cataract module, optional.
* Optional cataract module. Example order; no exam time is published.

Topography

22 rings. The whole anterior surface.

Placido rings reflect on the cornea; their shape gives the curvature, point by point. One-touch comparative and differential maps, keratoconus probability index and CLMI: for ortho-k, contact lenses and the allergic child.

> 6,000
measured points
28–67.5 D
corneal power, to 0.01 D
Schematic. Placido sees the anterior corneal surface; it guides, tomography confirms.

Aberrometry, new on the PRO

Cornea or lens: what will remain after phaco.

The Hartmann-Shack sensor measures the wavefront of the whole eye. The MYAH PRO splits it into corneal and internal aberrations: what comes from the cornea will remain after surgery, what comes from the lens leaves with it. A help when choosing a premium IOL.

880 nm
Hartmann-Shack
8 mm
maximum analysed pupil
Hartmann-Shack spotsTotalCornealInternal=+
Schematic. Illustrative maps, no values.

Pupillometry

From photopic to scotopic.

From photopic to scotopic, static and dynamic: pupil size and centration, for ortho-k, contact lenses and IOL choice.

2–10 mm
diameter, to 0.01 mm
PhotopicScotopic

6,4 mm (example)

Schematic. Example value.

Ocular surface

A stable tear film before keratometry.

Harmattan, dust, air conditioning: the ocular surface is often compromised. Tear break-up time is measured without dye, in the same device as keratometry, before IOL calculation or a contact lens fitting.

0.5–30 s
non-invasive break-up, to 0.1 s
0.10–1 mm
tear meniscus

6,4 s

Schematic, in real time. Example value.

Meibography

Meibomian glands, graded and followed up.

Meibography with grading, fluorescein images and video, inter-blink interval and questionnaires, in line with TFOS DEWS III. A basic assessment; for a dedicated dry eye clinic, see the TERA.

1–100%
gland loss
Schematic. Three shortened glands.

Cataract module, optional

Biometry, and 13 formulas.

With the cataract module, the MYAH PRO also measures corneal thickness, anterior chamber depth and lens thickness, and calculates the IOL with the Barrett formulas (Universal II, Toric, True-K, Rx), Olsen, Haigis or Hoffer Q. EVO and KANE are announced.

0.001 mm
central corneal thickness*
13
IOL formulas*
Axial lengthCornea*IOL*13 formulas
Schematic. * Optional cataract module. In very dense cataracts, ultrasound may still be needed.

MYAH PRO, ALADDIN or TERA?

Three devices, three uses.

MYAH PRO

For
Myopia, contact lenses, full work-up in one device
Axial length
14–40 mm, 1,060 nm interferometry
Objective refraction
Yes, Hartmann-Shack
IOL calculation
13 formulas (optional module)
Ocular surface
Basic assessment
For
Biometry and topography before cataract surgery
Axial length
Low-coherence interferometry
Objective refraction
n/a
IOL calculation
Barrett and Olsen
Ocular surface
n/a

Learn more : ALADDIN

n/a: not stated in our sources. Sources: Topcon MYAH PRO brochure (08.26), Topcon 2026 catalogue, Topcon website.

Specifications.

Keratometry and topography
Topography
Placido, 22 rings; over 6,000 measured points
Coverage
0.5 to 9.8 mm (8 mm sphere)
Radius of curvature
5.00 to 12.00 mm, to 0.01 mm
Corneal power
28.00 to 67.50 D, to 0.01 D
White-to-white
8.00 to 15.00 mm, to 0.01 mm
Keratoconus
Probability index, CLMI
Refraction and aberrometry
Sensor
Hartmann-Shack wavefront, 880 nm
Sphere
-25.00 to +22.00 D, to 0.01 D
Cylinder
0 to ±10.00 D, to 0.01 D
Aberrations
Total, corneal and internal
Pupil
2.00 to 10.00 mm; photopic to scotopic
Pupillary distance
20 to 85 mm
Biometry
Principle
Low-coherence interferometry, 1,060 nm SLED
Axial length
14 to 40 mm, to 0.01 mm
Anterior chamber*
1.5 to 7.0 mm, to 0.01 mm
Lens*
1.5 to 6.5 mm, to 0.01 mm
Corneal thickness*
0.2 to 1.2 mm, to 0.001 mm
IOL formulas*
Barrett Universal II, Toric, True-K, Rx; Olsen, Haigis, Holladay 1, Hoffer Q, SRK II, SRK/T, Camellin-Calossi, Shammas, Abulafia-Koch
Ocular surface
Tear break-up
Non-invasive, 0.5 to 30.0 s, to 0.1 s
Tear meniscus
0.10 to 1.00 mm
Meibomian glands
Gland loss 1 to 100%, with grading
Other
Inter-blink interval, fluorescein, questionnaires
Device
Connectivity
DICOM (storage, query, worklist), XML export, Harmony
Dimensions
564 × 332 × 598 mm (D × W × H)
Weight
21 kg
Operating conditions
10 to 40 °C; 8 to 75% humidity, non-condensing
Class
Class IIa medical device
Manufacturer
Visia Imaging Srl (Topcon group)

* Optional cataract module. Source: Topcon MYAH PRO brochure (EM000086, 08.26). Availability varies by country; specifications subject to change.

  1. Kobia-Acquah et al., 2022 (meta-analysis, childhood myopia in Africa).
  2. Wolffsohn et al., TFOS DEWS III, 2025.

In detail.

Following myopia, and much more

The MYAH PRO measures axial length by low-coherence interferometry (1,060 nm, 14 to 40 mm) and plots it on age-based reference curves. It also performs objective refraction and aberrometry (Hartmann-Shack), 22-ring Placido topography, pupillometry and a basic ocular surface assessment. With the optional cataract module, it measures the anterior segment and calculates the IOL.

Class IIa medical device, manufactured by Visia Imaging Srl (Topcon group). For healthcare professionals only; training is required before use. Availability varies by country.

Installation, training and service

Emetrop Afrique delivers, installs and commissions the device, trains the users and maintains it in the countries we cover. The device operates between 10 and 40 °C: we check the room’s air conditioning before installation.

Frequently asked questions.

What does the MYAH PRO measure?

Axial length, objective refraction, corneal topography, aberrations, the pupil and the ocular surface. With the optional cataract module: corneal thickness, anterior chamber depth, lens thickness and IOL calculation.

How does it measure axial length, and over what range?

By low-coherence interferometry (1,060 nm), from 14 to 40 mm, to a hundredth of a millimetre.

What are the myopia follow-up curves for?

To place a child's axial length against age-based reference data and follow it from one visit to the next. They help follow-up and explanations to parents; they do not replace the clinical examination.

How does it differ from the MYAH?

The MYAH PRO adds objective refraction and Hartmann-Shack aberrometry (total, corneal and internal), axial length from 14 to 40 mm and, as an option, anterior segment biometry and IOL calculation.

Can it be used for IOL calculation?

Yes, with the optional cataract module: 13 formulas, including Barrett Universal II, Toric, True-K and Rx, Olsen, Haigis and Hoffer Q. EVO and KANE are announced. In very dense cataracts, ultrasound may still be needed.

Can a technician or orthoptist use it?

Yes, it is designed for delegated testing, with alignment guidance. Training is required before use; we provide it at installation.

Does it connect to our IT systems?

Yes: DICOM (storage, query, worklist), XML export to the patient record, Harmony, and networked review software.

Images: Topcon Healthcare · Diagrams Emetrop Afrique

See the MYAH PRO at your site?

We bring it and show it to your team. No commitment.