TRK-3 OMNIA

Topcon · 4-in-1 pre-test device

Topcon TRK-3 OMNIA : TRK-3 OMNIA

Four measurements. One setup.

Topcon TRK-3 OMNIA – 4-in-1 pre-test device

Highlights.

Space

The space of one device. The work of two.

According to Topcon, it replaces the refractometer and tonometer in the space of one. It fits against a wall, in a corner or in the centre, and is operated from behind, from the side or face to face.

AutorefractorTonometerTRK-3 OMNIASpace freed
Schematic, top view.

Pressure

A thin cornea underestimates IOP. The TRK-3 also measures pachymetry.

It measures corneal thickness and adjusts the pressure with a formula the user can set. A screening tool: diagnosis uses applanation and visual field.

7–60 mmHg
tonometry
0.380–0.750 mm
pachymetry
thickness15 · measured18 · adjustedmmHg
Example values, from the Topcon brochure table (thin cornea). Schematic.

Small pupil, cloudy media

Small pupil. Hazy lens. The measurement gets through.

An SLD source and a Topcon-patented rotary prism: the measurement passes through a 2 mm pupil and, according to Topcon, reduces acquisition problems in cloudy media (confirmed on test eyes).

Rotary prismØ 2 mmRetina
Schematic. A dense cataract can make any autorefraction impossible.

Range

From -30 to +25 dioptres.

The widest sphere range in our Topcon range.

  • MYAH PRO−25 to +22 D
  • TRK-3 OMNIA−30 to +25 D
Sphere, Topcon brochures (TRK-3 OMNIA, MYAH PRO 08.26). Comparison within the Topcon range.

The refraction chain

Measured once. Never re-typed.

The autorefractor and lensmeter send their measurements to the CV-5000 refractor, which drives the test chart. According to Topcon, the transfer avoids transcription errors.

AutorefractorLensmeterRefractorTest chartSph · Cyl · AxisPrevious glassesE F P
Schematic. LAN or RS-232C links; only "some non-Topcon devices".

Specifications.

Refraction and keratometry
Sphere
-30 to +25 D (0.12 or 0.25 D steps)
Cylinder
0 – ±12 D
Minimum pupil
Ø 2.0 mm
Keratometry
Radius 5.00 to 13.00 mm
Corneal diameter
2.00 to 14.00 mm
Tonometry and pachymetry
IOP
7 to 60 mmHg
Pachymetry
0.380 to 0.750 mm
Adjustment
IOP adjusted to pachymetry, user-set formula
Device
Weight
23.2 kg
Power
100–240 V, 50–60 Hz, 80 VA
Connectivity
CV-5000, IMAGEnet®, DICOM, EMR
Class
IIa

Source: Topcon TRK-3 OMNIA brochure (2024). Specifications subject to change.

In detail.

Autorefkeratometer, tonometer and pachymeter

The TRK-3 OMNIA measures refraction, curvature, pressure and corneal thickness in one setup. It prepares the consultation: the doctor validates and prescribes.

Installation, training and follow-up

Emetrop Afrique supplies it, installs it, trains teams and provides follow-up across West and Central Africa.

Frequently asked questions.

Does it replace Goldmann tonometry?

No. The air tonometer is for pre-testing and screening; glaucoma diagnosis and follow-up use applanation, visual field and OCT.

Is it suitable for children?

It measures children, but without cycloplegia autorefraction overestimates myopia: cycloplegic refraction remains the reference before age 12.

Who installs it and trains the team?

Emetrop Afrique: installation, handover, training, then follow-up.

How much does it cost?

It depends on the configuration. Ask for a quote: it is free and without commitment.

Images: Topcon Healthcare · Diagrams Emetrop Afrique

See the TRK-3 OMNIA at your site?

We bring it and show it to your team. No commitment.