Skip to content

TERA

Topcon · dry eye imager

Topcon TERA : TERA

Dry eye, layer by layer.

Topcon TERA – dry eye imager

Highlights.

One work-up, every cause

The MYAH PRO spots it. The TERA finds the cause.

TFOS DEWS III no longer stops at the diagnosis: it asks what is drying the eye, to choose the treatment. The TERA brings tear film, lids and surface together on one device, each with its scale.

4
recognised scales
OSDI-6
and DEQ-5, built in
  • Questionnaire (OSDI-6)Moderate symptoms
  • Stability (break-up)Unstable
  • Volume (meniscus)Normal
  • Lipids (Guillon)2 · Wave
  • Glands (Pult)Grade 2
  • Lid margin (Efron)Grade 1
  • Redness (Efron)Grade 1
  • Staining (Jenvis)Grade 0
Example report. Grades are assigned by the clinician; colours aid reading, the diagnosis remains the doctor's.

Break-up time

Non-invasive break-up, sector by sector.

The patient blinks, then keeps the eye open. The TERA tracks the red rings reflected on the cornea and records when and where the film breaks. Meanwhile it also films the curvature map: you see keratometry shift with the tear film.

0.5–30 s
to 0.1 s, no dye
< 10 s
TFOS DEWS III cut-off

5,8 s

Schematic, in real time. Example value.

Blinking

When the film breaks before the next blink.

Between two blinks the film holds, then breaks: if it breaks before the next blink, the eye is left exposed. The TERA measures the inter-blink interval and derives the ocular protection index.

0 s5 s10 s15 s20 sBlinkStable filmEye exposedOPI = break-up time ÷ inter-blink intervalTime →
Schematic. Example blinks and break-ups.

Lipid layer

The lipid layer, graded on the Guillon scale.

Under diffuse light, the lipid film shows fringes and patterns. The TERA photographs or films it; the clinician grades it on Dr Guillon's scale. A poor layer points to evaporative dry eye, common with air conditioning and screens.

Guillon
5-grade scale
0 · Colour fringes1 · Amorphous2 · Wave3 · Closed meshwork4 · Open meshworkGrade assigned by the clinician
Schematic. Qualitative assessment, no thickness measurement.

Meibomian glands

The glands, at 940 nm.

With the lid everted, infrared light reveals the glands. The clinician outlines the area; the TERA computes the loss area, graded on Pult's Meiboscale, upper and lower lid.

1–100%
loss area
940 nm
infrared
Schematic. Three shortened glands.

Tear meniscus

Tear meniscus height, in three values.

At the lower lid margin, the operator places the cursors; the software traces the meniscus profile and gives the minimum, mean and maximum height.

0.10–1 mm
to 0.01 mm
0.18 mm
minimum
0.24 mm
mean
0.31 mm
maximum
Schematic. Cursors placed by the operator, profile computed by the software. Example values.

Before cataract surgery

Treat the surface, then calculate the IOL.

An unstable tear film makes keratometry vary, and with it the IOL calculation. In a US study, 62.9% of cataract candidates had a break-up time of 5 s or less. Treat the surface, measure again, then calculate the IOL: the TERA prepares biometry, it does not replace it.

62.9%
break-up ≤ 5 s before cataract (Trattler 2017)
± 0.12 D
keratometry repeatability, per the manufacturer

A delegated clinic

Your team measures. You decide.

Topography, pupil and break-up time are aligned and captured automatically; the rest uses the joystick or touchscreen. You review, grade and print from your workstation.

TERA Review
review on any networked PC
  1. Questionnaire
  2. Images and measurements
  3. Grading by the doctor
  4. Report for the patient
  5. Follow-up visit to visit
Example order; no exam time is published.

TERA, MYAH PRO or ALADDIN?

Three devices, three uses.

TERA

For
Dry eye clinic, adults
Dry eye
Full work-up, 4 scales
Lipid layer, redness, blepharitis
Yes
Topography
Placido, 22 + 2 rings
Axial length
No
For
Myopia, contact lenses, full work-up in one device
Dry eye
Basic assessment
Lipid layer, redness, blepharitis
No
Topography
Placido, 22 rings
Axial length
14–40 mm

Learn more : MYAH PRO

n/a: not stated in our sources. Sources: Topcon TERA brochures and user manual (2025), Topcon MYAH PRO brochure (08.26), Topcon 2026 catalogue.

Specifications.

Ocular surface
Tear break-up
Non-invasive, 0.5 to 30.0 s, to 0.1 s; sector map
Blinking
Interval, rate; ocular protection index (OPI)
Tear meniscus
0.10 to 1.00 mm, to 0.01 mm
Lipid layer
Diffuse light, photo or video; Guillon scale
Meibomian glands
Infrared 940 nm; loss area 1 to 100%; Pult Meiboscale
Redness
Bulbar and limbal; Efron and Jenvis scales
Blepharitis
Lid margin in colour; Efron scale
Staining
Fluorescein (image and video), lissamine green
Questionnaires
OSDI-6, DEQ-5
Cornea and pupil
Topography
Placido, 22 + 2 rings; over 6,000 measured points; up to 9.5 mm
Keratometry
28.00 to 67.50 D, to 0.01 D; radius 5.00 to 12.00 mm
Keratoconus
Probability index, CLMI
Aberrations
Anterior corneal Zernike
Pupil
2.00 to 10.00 mm; photopic to scotopic
White-to-white
8.00 to 15.00 mm
Device
Automatic alignment
Topography, pupil, break-up time
Review
TERA Review on any networked PC; PDF and XML export
Dimensions
332 × 555 × 552 mm (W × H × D)
Weight
19 kg
Operating conditions
10 to 40 °C; 8 to 75% humidity, non-condensing
Class
Class IIa medical device (MDR), adult patients
Manufacturer
Visia Imaging Srl (Topcon group)

Sources: Topcon TERA brochures (2025), user manual (Rev. 1.0, 2025), declaration of conformity. Availability varies by country; specifications subject to change.

  1. Wolffsohn et al., TFOS DEWS III, 2025; Trattler et al., 2017 (PHACO study).

In detail.

The dry eye work-up, cause by cause

The TERA brings together the tear film (stability, volume, lipid layer), the lids (meibomian glands, lid margin, blinking) and the surface (redness, staining), with the patient’s questionnaires and corneal topography. Each test comes with its reference scale; the clinician assigns the grades and makes the diagnosis.

Class IIa medical device (Regulation (EU) 2017/745), manufactured by Visia Imaging Srl (Topcon group). For healthcare professionals, for adult patients; training is required before use. Availability varies by country.

Installation, training and service

Emetrop Afrique delivers, installs and commissions the device, trains the users and maintains it in the countries we cover. The device operates between 10 and 40 °C: we check the room’s air conditioning before installation.

Frequently asked questions.

What does the TERA measure?

Tear film stability without dye, blinking, tear meniscus height, the lipid layer, meibomian glands, redness, the lid margin and staining (fluorescein, lissamine green). It includes the OSDI-6 and DEQ-5 questionnaires and also performs corneal topography and pupillometry.

Does the TERA follow TFOS DEWS III?

Topcon presents it as aligned with TFOS DEWS III (2025): OSDI-6 questionnaire, non-invasive break-up time, staining and tests that help find the cause (lipids, tear volume, lids). Tear osmolarity is not among its measurements.

Does the TERA make the diagnosis?

No. It measures, images and helps grade with recognised scales; the clinician assigns the grades and makes the diagnosis with the clinical examination.

How does it differ from the MYAH PRO's dry eye assessment?

The MYAH PRO offers a basic assessment (break-up time, meniscus, glands, fluorescein, questionnaires) within a myopia management device. The TERA adds the lipid layer, redness, blepharitis, lissamine green and grading scales: it is the dry eye clinic device.

Can a technician perform the exam?

Yes, after training. Topography, pupil and break-up time are aligned and captured automatically; other tests use the joystick or touchscreen. The doctor reviews and grades the images from their own workstation.

Can it be used before cataract or refractive surgery?

Yes: Topcon intends it for ocular surface assessment before and after surgery. It does not measure axial length; it complements the biometer.

Does the TERA suit our climate?

It operates from 10 to 40 °C and 8 to 75% non-condensing humidity: an air-conditioned room is required. We install it, train the team and service it.

Images: Topcon Healthcare · Diagrams Emetrop Afrique

See the TERA at your site?

We bring it and show it to your team. No commitment.